The Model 2930 can also be used with hexane based EPA method 1664, which is an analysis technique that does not require the use of freon as as solvent. A special cell is an optional accessory for this freon-free technique.
N/A
Water and hydrocarbons absorb in the near IR at 3.4 to 3.5 microns (2930 cm-1). By dissolving the oil content of a sample into fluorochlorocarbon (freon), which is nonabsorbing at 3.4 to 3.5 microns, and following the extraction procedure established in EPA methods 413.2 and 418.1, a working curve is prepared using blank solvent (freon) and known standards. Samples are then compared to the calibration curve in direct ppm readings shown on the instrument. Repeatability is ±1% on identical samples. |