The ICL Hydrocarbon Test Kit contains a complete set of slide-mounting accessories for oil-in-water analysis in all concentrations that can be used in any FTIR, dispersive or scanning spectrophotometer. This kit will permit the user to test for the presence of hydrocarbons in water in any concentration using EPA methods 413.2 or 418.1. It consists of matched pairs of 10mm, 50mm and 100mm IR Grade Quartz cylindrical cells and holders for each cell size. The cylindrical cells are matched to a standard so that if one of the cells in the set is broken, it can be replaced with another matched cell without the need to replace the pair. Matched cell pairs allow precise repeat measurement of the same sample. The kit is packed in a durable portable carrying case that protects the cells and makes them easy to transport without fear of breakage.
N/A
Water and hydrocarbons absorb in the near IR at 3.4 to 3.5 microns (2930 cm-1). By dissolving the oil content of a sample into fluorochlorocarbon (freon), which is nonabsorbing at 3.4 to 3.5 microns, and following the extraction procedure established in EPA methods 413.2 and 418.1, a working curve is prepared using blank solvent (freon) and known standards. Samples are then compared to the calibration curve in direct ppm readings shown on the instrument. Repeatability is ±1% on identical samples. |